Introduction
Multifunction process calibrator | DTE-35G is a multi-function, high-precision portable instrument. It is specially designed for process control, suitable for industrial field debugging, can meet a wide range of work requirements, support the measurement and output of almost all process parameters such as voltage, resistance, current, frequency, thermocouple, thermal resistance The secondary instruments of these.
DTE-35G multifunction calibrator are calibrated accuracy can reach 0 . 02%.
Function
◇ Super protection function: waterproof grade IP67, 220V automatic protection when any signal terminal is wrongly connected;
◇ Measurement / output: voltage, current, frequency, resistance, of which current output supports active and passive;
◇ Simulate thermal resistance and thermocouple in the form of temperature;
◇ Both thermocouple measurement and output provide automatic and manual cold junction temperature compensation;
◇ New thermocouple types can be customized according to customer needs;
◇ Simulate two-wire transmitter;
◇ Accuracy is 0 . 02%, 0 . 05%;
◇ Two isolated channels support simultaneous measurement and output;
◇ Can provide manual stepping, automatic stepping, 0 ~ 100% step and slope output function;
◇ With white LED backlight, with manual backlight adjustment and automatic power off function, suitable for on-site use;
◇ Provide DC24V loop power supply for on-site debugging;
◇ Unique automatic identification of three-wire and four-wire wiring;
◇ Adopt standard 7 # battery, which is convenient for customers to replace and purchase;
◇ APSL high-precision intelligent pressure module can be selected according to customer needs to achieve pressure measurement function;
Technical parameters
Measurement standard:
DTE-35G multifunction calibrator Features | DTE-35G process calibrator Range | Accuracy | |
DTE-35G | DTE-25G | ||
Voltage | 0-60.000V(On screen±30V) | 0.02%+2 | 0.05%+2 |
Electric current | 0~24.000mA(On screen±24mA) | 0.02%+2 | 0.05%+2 |
Millivolt | -15.000mV~80.000mV | 0.02%+2 | 0.05%+2 |
80.00mV~125.00mV | |||
Resistance (2-wire 3-wire) | 0.00Ω~440.00Ω | 0.15Ω | 0.25Ω |
400.0Ω~3200.0Ω | 1.0Ω | 1.5Ω | |
Resistance (4-wire) | 0.00Ω~440.00Ω | 0.10Ω | 0.15Ω |
400.0Ω~3200.0Ω | 0.5Ω | 1.0Ω | |
Frequency | 1.000Hz~99.999Hz | 0.01%+1 | 0.02%+1 |
100.00Hz~999.99Hz | |||
1000.0Hz~9999.9Hz | |||
10.000kHz~99.999kHz | |||
Thermocouple | J、K、T、E、R、S、B、N | ||
Thermal resistance | Pt100、Pt1000、Cu50、Cu100 |
Output indicators:
DTE-35G multifunction calibrator Features | DTE-35G process calibrator Range | Accuracy | |
DTE-35G | DTE-25G | ||
Electric current | 0~10.000V | 0.02%+2 | 0.05%+2 |
Active current | 0~24.000mA | 0.02%+2 | 0.05%+2 |
Passive current | 0~24.000mA | 0.02%+2 | 0.05%+2 |
Millivolt | -15.000mV~99..999mV | 0.02%+2 | 0.05%+2 |
100.00mV~125.00mV | |||
Resistance | 10.00Ω~440.00Ω | 0.15Ω | 0.25Ω |
400.0Ω~3200.0Ω | 1.0Ω | 1.5Ω | |
Frequency | 0.20Hz~200.00Hz | 0.01%+1 | 0.02%+1 |
200.0Hz~200.00Hz | |||
2.000kHz~20.000kHz | |||
Thermocouple | J、K、T、E、R、S、B、N | ||
Thermal resistance | Pt100、Pt1000、Cu50、Cu100 |
Other parameters:
Operating temperature: -10 ℃ ~ 55 ℃
Storage temperature: -20 ℃ ~ 60 ℃
Weight: 650g
Dimensions: 200mm * 96mm * 47mm
Power supply: 6 pieces of 7 # nickel-metal hydride batteries or AC adapter
Standard accessories:
Table pen: 18A53 table pen 2 pay
Power adapter: one 18A51 charger.
Optional accessories:
RS232 communication line: 18P01 communication line
Annex:
Previous:DTY2005 intelligent pressure calibration device | next:Custom temperature and humidity verification/calibration chamber |
Headquarter Address :Tai'an High-tech Development Zone, Shandong Province
Telephon :+86 0538-5089056 +86 13605384645 +86 13953856217
Fax : +86 0538-5059718
After sale:+86 0538-5050959 +86 13605384645
Website : www.nenoglobal.com
E-mail : tadt17@dearto.cn / tadtzdh@dearto.cn
Copyright© Tai'an Detu Automation Instrument Co., Ltd. Lu ICP 14024757-1